EPA Method 360.1


ALS - Columbia offers EPA Method 360.1 testing at these laboratories:

Jacksonville Jacksonville, Florida Laboratory
Simi Valley Simi Valley, California Laboratory

Method information displayed is provided for informational purposes only. No warranty (express or implied) is made as to the website accuracy, completeness, or applicability (such as the age of a method and whether or not it applies to your project). Please contact us for assistance. Analytical testing dots

PDF IconView Actual EPA Method 360.1 (PDF File)


EPA Method 360.1:
Dissolve Oxygen by Membrane Electrode. Official Name: Oxygen, Dissolved (Membrane Electrode)

Summary:
Dissolved oxygen in a sample is measured using a probe. The most common probes depend on electrochemical reactions. Under steady state conditions, the current or potential is correlated to dissolved oxygen concentrations.

Scope:
This method determines dissolved oxygen in water and is recommended as an alternative to the modified Winkler procedure when interferences are present, a non-destructive method is desired, or when continuous monitoring is desired.

Citation:
Methods for the Chemical Analysis of Water and Wastes (MCAWW) (EPA/600/4-79/020)

Interferences:
(A) Inorganic salts: Salts can change probe response. Use conversion factors to correct for response in saline waters.(B) Reactive gases: Reactive gases like chlorine and hydrogen sulfide can permeate the probe and disrupt operation.Temperature sensitivity: Temperature can effect a probe's output. Generally, a manufacturer provides a compensation factor.

QC Requirements:
None.

Maximum Holding Time:
Analyze immediately (MCAWW, Table 1).

Media:
WATER

Subcategory:
Inorganic

Concentration:
The range can vary depending on the instrument. 0-20 mg/L is given as an example.

Sample Prep:
None.

Precision:
Precision and accuracy data are not available.

Detection:
Generally, meters have a sensitivity of 0.05 mg/L or better.

Revision Number:
Issued 1971

Test Description:
Oxygen, Dissolved -

Report Number:
600/4-79-020

Instrument used for this test:
DOMeter