Method information displayed is provided for informational purposes only. No warranty (express or implied) is made as to the website accuracy, completeness, or applicability (such as the age of a method and whether or not it applies to your project). Please contact us for assistance. Analytical testing dots

PDF IconView Actual EPA Method 365.1 (PDF File)


EPA Method 365.1:
Phosphorus (all forms) by Semi-Automated Colorimetry. Official Name: Phosphorus, All Forms (Colorimetric, Automated, Ascorbic Acid)

Summary:
A sample is appropriately treated to convert all phosphorus of interest to reactive orthophosphate. Ammonium molybdate and antimony potassium tartrate are added to the treated sample reacting with orthophosphate in an acidic medium to form an antimony-phospho-molybdate complex. This complex is reduced to an intensely blue-colored complex by ascorbic acid. The concentration of the orthophosphate is measured by detecting the absorbance of the complex with a spectrophotometer.Note: In Section 7.7 of the method (preparation of 11 N sulfuric acid solution), "600 mL reagent water" should read "300 mL reagent water."

Scope:
This method determines total phosphorus, hydrolyzable phosphorus, and orthophosphate in drinking, ground, and surface waters; and domestic and industrial wastes.

Citation:
Methods for the Determination of Inorganic Substances in Environmental Samples (EPA/600/R-93/100)

Interferences:
(A) Metals and silica: Copper, iron, and silica do not interfere at the levels reported in sea water, but excessively high concentrations of iron can cause precipitation and loss of phosphorus.(B) Salt error: Salt error for 5%-20% salt samples was less than 1%.(C) Arsenate: Arsenate can cause a positive interference, but is often at low concentrations.(D)Turbidity: Remove turbidity via filtration.

QC Requirements:
The minimum recommended quality control requirements include an initial demonstration of laboratory capability (determining the linear calibration range and method detection limit, and analyzing a quality control sample; LCR and MDL, and QCS), and the periodic analysis of laboratory reagent blanks.

Maximum Holding Time:
48 hours (orthophosphate, dissolved)28 days (Hydrolyzable and Total)24 hours (Total Dissolved).(MCAWW, Table 1).

Media:
WATER

Subcategory:
Inorganic

Concentration:
0.01 - 1.0 mg/L

Precision:
Precision and accuracy values were calculated using interlaboratory data from EPA-managed Water Pollution (WP) performance evaluation studies.

Detection:
MDL procedure is equivalent to the procedure at 40 CFR part 136, Appendix B.

Revision Number:
Revision 2.0, August 1993

Test Description:
Orthophosphate (as P

Report Number:
600/R-93-100

Instrument used for this test:
Discrete-FIA

Analytical testing dots
At least 2 analytes can be tested by ALS - Columbia using EPA Method 365.1. Please contact us for all available analytes which can be tested by EPA Method 365.1.

AnalyteCAS Number
Orthophosphate as Phosphorus14265-44-2A
Phosphorus7723-14-0





Examples:  8260
TO-15

Examples:  Dioxin
Mercury
Analytical testing dots


<-- Search All Test Methods



Suggestions? The test methods page continues to expand and improve. If you have suggestions for improvement, we would enjoy hearing from you. Please contact the webmaster here.


Analytical testing dots