EPA Method 375.1


EPA Method 375.1 does not directly match ALS - Columbia's list of currently available methods. Please call for availability.


Method information displayed is provided for informational purposes only. No warranty (express or implied) is made as to the website accuracy, completeness, or applicability (such as the age of a method and whether or not it applies to your project). Please contact us for assistance. Analytical testing dots

PDF IconView Actual EPA Method 375.1 (PDF File)


EPA Method 375.1:
Sulfate by Colorimetry. Official Name: Sulfate (Colorimetric, Automated, Chloranilate)

Summary:
Sulfate in a sample is treated with barium chloranilate to precipitate barium sulfate and release a highly colored chloranilate ion. The absorbance of the chloranilic acid is measured with a colorimeter and is proportional to sulfate concentration.

Scope:
This automated method determines sulfate in drinking and surface waters; domestic and industrial wastes.

Citation:
Methods for the Chemical Analysis of Water and Wastes (MCAWW) (EPA/600/4-79/020)

Interferences:
Cations: Cations such as calcium, aluminum, and iron interfere by precipitating the chloranilate. These cations can be removed by passing the sample through a ion exchange column.

QC Requirements:
None.

Maximum Holding Time:
28 Days (MCAWW, Table 1).

Media:
WATER

Subcategory:
Inorganic

Concentration:
10 to 400 mg sulfate/L

Sample Prep:
None.

Precision:
Precision and accuracy values were calculated using interlaboratory data from EPA-managed Water Pollution (WP) performance evaluation studies.

Detection:
None.

Revision Number:
Issued 1971