EPA Method 8021B
ALS - Columbia offers EPA Method 8021B testing at these laboratories:
![]() |
Kelso, Washington Laboratory |
![]() |
Rochester, New York Laboratory |
![]() |
Simi Valley, California Laboratory |
Method information displayed is provided for informational purposes only. No warranty (express or implied) is made as to the website accuracy, completeness, or applicability (such as the age of a method and whether or not it applies to your project). Please contact us for assistance.

View Actual EPA Method 8021B (PDF File)
EPA Method 8021B:
Aromatic and Halogenated Volatiles by Gas Chromatography Using Photoionization and/or Electrolytic Conductivity Detectors
Summary:
Samples are analyzed using either direct injection, purge-and-trap, or vacuum distillation. The compounds are separated by the GC and detected using either a photoionization detector and an electrolytic conductivity detector in series or by using separate detection.
Scope:
This method determines the concentration of various volatile organic compounds in ground water, aqueous sludges, caustic liquors, waste solvents, oily wastes, mousses, tars, fibrous wastes, polymeric emulsions, filter cakes, spent carbons, spent catalysts, soils, and sediments.
Citation:
SW-846 Online: Test Methods for Evaluating Solid Waste, Physical/Chemical Methods
Interferences:
(A) Contamination: Sample contamination may occur by diffusion of volatile organics (particularly chlorofluorocarbons and methylene chloride) through the septum seal during shipment and storage. The analysis of trip blanks will identify this problem.(B) Matrix interferences: Sulfur dioxide may interfere in the analysis of vinyl chloride.(C) Refer to the appropriate 5000 Series method and Method 8000.
QC Requirements:
Refer to Chapter One and Methods 8000, 3500, and 5000 for specific QC procedures. Procedures include: validation of sample preparation, introduction, and analysis, initial demonstration of capability, calibration verification, evaluation of retention times, analysis of method blanks, matrix spikes, duplicates, surrogates and laboratory control samples.
Maximum Holding Time:
See SW-846 Chapter Four for information.
Media:
VARIOUS
Subcategory:
Organic
Concentration:
0.1 - 200 ug/L.
Sample Prep:
Methods 5035, 3585, 5030, 5021, 5032
Precision:
Precision and accuracy values were calculated based on the average of 7 replicates, spiked at 10 ug/L, using internal calibration, the purge-and-trap method, and hall electrolytic conductivity detection.
Detection:
None.
Revision Number:
Revision 2, December 1996
Test Description:
Halogenated Volatile
Report Number:
SW-846 Ch 4.3.1
Instrument used for this test:
GC
At least 85 analytes can be tested by ALS - Columbia using EPA Method 8021B. Please contact us for all available analytes which can be tested by EPA Method 8021B.
Examples: | 8260 TO-15 | |
| Examples: | Dioxin Mercury |