ALS - Columbia may or may not offer methods shown. Please contact us to learn about ALS - Columbia's latest available methods. No warranty (express or implied) is made as to the website accuracy, completeness, or applicability (such as the age of a method and whether or not it applies to your project).

Analytical testing dots

PDF IconView Actual USGS Method I-2462 (PDF File)


Source:
USGS

Method Name:
Mercury, dissolved, CVFAA. Official Name: Mercury, atomic absorption spectrometric, cold vapor, flameless automated-sequential

Description:
The cold-vapor, flameless, atomic absorption procedure is based on the absorption of radiation at 253.7 nm by mercury vapor. Organic mercury compounds, if present, are decomposed by hot (95oC) digestion with potassium dichromate and potassium persulfate in acid solution. Mercuric ions are then reduced to the elemental state with stannous chloride, and mercury vapor is subsequently removed from solution by aeration and passed through a cell positioned in the light path of an atomic absorption spectrometer. The method is based on a procedure described by El-Awady and others (1976).

Scope:
This method may be used to analyze water and wastewater containing at least 0.1 ug/L mercury. Samples containing mercury concentrations greater than 8.0 ug/L need to be diluted.

Citation:
Methods for the Determination of Inorganic Substances in Water and Fluvial Sediments, Techniques of Water-Resources Investigations of the United States Geological Survey, Book 5, Chapter A1 Edited by Marvin J. Fishman and Linda C. Friedman

Interferences:
Chloride concentrations up to 5,000 mg/L do not interfere; higher concentrations were not tested.Hydroxylamine hydrochloride-sodium chloride solution is added to prevent interference from residual chlorine.El-Awady and others (1976) reported that copper sulfate (1,000 mg/L) does not interfere and that chemical-oxygen-demand (COD) concentrations of less than 700 mg/L can be tolerated. Ethyl alcohol, methyl alcohol, glycerol, chloroform, and carbon tetrachloride did not interfere when added in concentrations as high as 0.5 percent. Major interferences were observed from benzene and toluene. A maximum tolerance of 500 ug/L was obtained for these compounds. Selenate concentrations up to 10,000 ug/L do not interfere; higher concentrations were not tested. Concentrations of selenite greater than 100 ug/L interfere by suppressing the mercury absorption.

QC Requirements:
Calibrate instrument using calibration standards (CAL). Quality control samples (QCS) and laboratory blanks (LB) analyzed at a minimum of I each after every 10 samples

Maximum Holding Time:
180 days

Media:
WATER

Subcategory:
Inorganic

Concentration:
0.1 ug/L to 8.0 ug/L mercury

Precision:
-

Detection:
-

Revision Number:
1985

Analytical testing dots

Related Methods:
Test Method 500001    Test Method 500007    Test Method 500021    Test Method 500061    Test Method 500071    Test Method 500076    Test Method 500081    Test Method 500086    Test Method 500090    Test Method 500201   


<-- Search again



*Method information displayed is provided as a convenience only. No warranty (express or implied) is made as to its accuracy, completeness, or applicability (such as the age of a method and whether or not it applies to your project).

Suggestions? The test methods page continues to expand and improve. If you have suggestions for improvement, we would enjoy hearing from you. Please contact the webmaster here.


Analytical testing dots