ALS - Columbia may or may not offer methods shown. Please contact us to learn about ALS - Columbia's latest available methods. No warranty (express or implied) is made as to the website accuracy, completeness, or applicability (such as the age of a method and whether or not it applies to your project).

Analytical testing dots

PDF IconView Actual USGS Method I-2477-92 (PDF File)


Source:
USGS

Method Name:
Determination of metals in Water by ICP-MS. Official Name: Methods of Analysis by the U.S. Geological Survey National Water Quality Laboratory - Determination of Metals in Water by ICP-MS

Description:
Sample solution is pumped by a peristaltic pump into a pneumatic nebulizer, which generates a liquid aerosol. This aerosol is transported by argon gas flow into a water-cooled spray chamber where the large droplets are removed by gravity and condensation. The small droplets are further swept into a radiofrequency ICP, where evaporation, molecular dissociation, atomization, and ionization occur. The ions are physically extracted from the center of the plasma by a differentially pumped vacuum system, through a water-cooled sampler cone and skimmer cone assembly, which is in physical contact with the horizontally mounted ICP.The extracted ions are focused by an electrostatic ion lens assembly and accelerated in to a unit-resolution quadrupole MS. For a given combination of radiofrequency and direct current voltages applied to the quadrupoles, only ions of a specific mass-to-charge ratio pass through the quadrupoles and reach the detector. Rapidly scanning the voltages on the quadrupoles has the effect of rapidly scanning the mass spectrum at the detector. Selected mass values, where high ion signals are expected, can be pre-programmed to be skipped in the mass scan to avoid overload damage to the detector. The mass spectrometer alternatively can be operated in a peak-jumping mode, which rapidly changes the applied voltages to effectively jump to preselected masses instead of operating in the continuous mass-scanning mode already described.Mass-scanning is suggested in this method because the recorded spectrum can be recalled and qualitatively examined for the presence or absence of any analytes or potential interferents.The ion signal is detected by a continuous dynode electron multiplier of the channeltron type. The physical impact of an ion on the detector surface produces a pulse of electrons. The resulting signal is processed in a digital, pulse-counting mode. New or upgraded instrumentation with an "extended dynamic range" option allows operation of the detector in an analog mode for high analyte concentrations. In either case, the ion signal is electronically amplified, and the resulting data are processed by a multichannel analyzer, or alternative data system, and computer.

Scope:
This method is suitable for the single-element or multi-element determination of dissolved concentrations of 22 selected trace metals in water. The method is applicable to surface-water, ground-water, drinking-water, and precipitation samples that have a measured specific conductance of less than 2,500 uS/cm at 25oC. The method is applicable to metals in the concentration range from 1.0 to 1,000 ug/L. The use of new or upgraded instrumentation with an "extended dynamic range" option can extend the upper concentration limit to 200 mg/L. Also samples with a specific conductance greater than the specified limit may be analyzed after appropriate dilution to conform to the specified limit; however, method reporting limits for the original sample will be increased according to the dilution factor. The determination of dissolved arsenic, boron, lithium, selenium, strontium, thallium, and vanadium in filtered, acidified natural water was later added to the original method. While originally published as two stand-alone reports, the two have been combined into a single file for the benefit of NEMI users.

Citation:
Garbarino, J.R., 1999, Methods of analysis by the U.S. Geological Survey National Water Quality Laboratory--Determination of dissolved arsenic, boron, lithium, selenium, strontium, thallium, and vanadium using inductively coupled plasma-mass spectrometry: U.S. Geological Survey Open-File Report 99-093.

Interferences:
Physical interferences are associated primarily with sample introduction and are minimized by using the internal standardization technique. Isotopes measured in this procedure have been selected specifically to minimize spectral interferences from isobaric, doubly charged, and molecular ions. Multiple isotopes can be measured for selected elements that have potential isobaric or molecular ion interference. The analyst must be aware of these interferences because they might be present with certain types of sample matrices. Memory effects related to sample transport are negligible for most elements that are normally present in natural water. Thallium and vanadium did not recover to reagent-blank intensity levels within the rinse period; thallium and vanadium intensities were 10 and 2 times greater than reagent-blank levels, respectively. Consequently, the analyst must review all analytical results to ensure that errors from carryover are minimized. Sample matrix composition could also affect the bias and variability of ICP-MS determinations. The use of internal standardization compensates for most matrix effects, however, some matrix interferences remain problematic. Matrix composition can suppress the ionization efficiency of the plasma and result in negatively biased elemental concentrations.

QC Requirements:
Quality-control samples area analyzed at a minimum of one in every ten samples. These QC samples include at least one of each of the following: blanks, quality control samples, third party check solutions, replicates, and spikes. Correlation coefficients for calibration curves must be at least 0.99. QC samples must fall within 1.5 standard deviations of the mean value. If all of the data-acceptance criteria in the SOPs are met, then the analytical data are acceptable.

Maximum Holding Time:
180 days from sampling

Media:
WATER

Subcategory:
Inorganic

Concentration:
0.08 - 1,000 (undiluted)

Precision:
Spike recovery percentages were determined for the new elements in matrices that are representative of reagent water, surface water, and ground water. Seven replicate recoveries at 5 to 10 times the MDL (the low-level spike) and 75 mg/L (the high-level spike) were determined in each matrix over a period of about 1 week. Average recoveries in the reagent-water matrix ranged from 93 to 105 percent for all elements except boron, whose low-level spike recovery was 77 percent. Recovery variability for the low-level spike ranged from 4 to 11 percent, depending on the element and matrix.

Detection:
Method detection limits were obtained by analyzing seven aliquots of a laboratory-prepared multi-element solution with concentrations of the metals near the expected detection limit of the method. The seven aliquots were analyzed consecutively in a single analysis procedure. The results were blank-subtracted. The standard deviation of each analyte for the seven replicate analyses was multiplied by the appropriate t-value for 99- percent confidence level to obtain the corresponding method detection limit. For the analytes added to the original method: short-term MDLs were calculated based on guidance in USEPA CFR40 and represent pooled averages on the basis of four MDLs determined on different days over several weeks.

Revision Number:
1999

Analytical testing dots

Related Methods:
Test Method 500001    Test Method 500007    Test Method 500021    Test Method 500061    Test Method 500071    Test Method 500076    Test Method 500081    Test Method 500086    Test Method 500090    Test Method 500201   


<-- Search again



*Method information displayed is provided as a convenience only. No warranty (express or implied) is made as to its accuracy, completeness, or applicability (such as the age of a method and whether or not it applies to your project).

Suggestions? The test methods page continues to expand and improve. If you have suggestions for improvement, we would enjoy hearing from you. Please contact the webmaster here.


Analytical testing dots